Table 3.

Reported Comparative Radiation Effective Dose (E 2007) for Limited, “Focused” or Small FOV CBCT Systems.

Dosea
Absolute Comparative

CBCT unit Ref. Technique Effective dosea (μSv) Digital panoramic equivalentb No. of days of annual per capita backgroundc

Kodak 9000 3D [30] Mx Post/Mx Ant/Mn Post/Mn Ant 9.8/5.3/38.3/21.7 .7/.4/2.7/1.6 1.2/.6/4.7/2.6
PreXion 3D [18] Standard/High Res 189/388 13.5/27.7 23/47
ProMax 3D [18] Small/Large 488/652 35/47 59/79
3D Accuitomo [23] Ant (4 × 4 cm/6 × 6 cm) 20/43 1.4/3.1 2.5/5.2
[24] Min (Mn PM)—Max (Mn 3rd Mol) 11–77 .8–5.5 2.5–5.2
[19] Mx (Ant/PM/Mol) 29/44/29 2/3.2/2 3.5/5.3/3.5
Mn (Ant/PM/Mol) 13/22/29 .9/1.6/2 1.6/2.7/3.5
[17] II/FPD Large/FPD Small 30/102/50 2.1/7.3/3.6 3.6/12.4/6
Veraview [23] Ant (4 × 4 cm/8 × 4 cm/pan + 4 × 4 cm) 31/40/30 2.2/2.9/2.1 3.8/4.9/3.6
[25] 4 × 4 cm 2.9 .2 .06

aUsing 2007 ICRP calculations.

bMedian of published effective dose for digital dental panoramic radiography  =  14 μSv.

cAnnual per capita  =  3.0 mSv (3,000 μSv) per annum.