TABLE 1.

Technique Parameter Options and Ranges for the Digital Mammography Systems Used in the American College of Radiology Imaging Network (ACRIN) Digital Mammographic Imaging Screening Trial (DMIST)

Manufacturer Model Target Filtration kVp Range mAs Range

Fischer Imaging SenoScan W-Re Al 20–45 (22–39) 18–52, 80–230 mA; (18–49, 80–217 mA)
Fujifilma 5000D (CR) Mo Mo (24–30) (18–417)
Mo Rh (26–32) (56–240)
Rh Rh (30–31) (88–247)
GE Healthcare Senographe 2000D Mo Mo 22–35 (22–33) 4–400 (21–387)
Mo Rh 22–40 (26–31) 4–400 (58–399)
Rh Rh 25–49 (26–32) 4–320 (28–317)
Hologic (CCD or Selenia) Fiber Optic-CCD Array Mo Mo 20–35 (24–32) 3–600 (35–489)

Note—For ranges, the first listing is the theoretic range provided by the manufacturer, the ranges in parentheses are ranges of parameters used in ACRIN DMIST. W-Re = tungsten-rhenium, Al = aluminum, Mo = molybdenum, Rh = rhodium.

a

Automatic exposure control was used on existing mammography systems with Fujifilm CR cassettes. Density control was adjusted to approximately equalize dose between screen-film mammography and CR.