Summary of adjusted risk ratios. Revision during the entire study period

Endpoint Population investigated Reference population Adjusted RR (CI) p-value
Revision of any component
 for any reason All uncemented implants All cemented implants 1.5 (1.4–1.6) < 0.001
 due to aseptic loosening All uncemented implants All cemented implants 1.5 (1.3–1.6) < 0.001
 due to deep infection All uncemented implants All cemented implants 0.9 (0.6–1.3) 0.5
 for any reason All uncemented implants from 2000 All cemented implants from 2000 1.5 (1.2–1.8) < 0.001
Cup revision due to aseptic loosening All uncemented cups All cemented cups 1.8 (1.6–2.0) < 0.001
Stem revision due to aseptic loosening All uncemented stems All cemented stems 0.4 (0.3–0.5) < 0.001
Revision for any reason within 2 years of the index procedure All uncemented implants All cemented implants 1.8 (1.5–2.2) < 0.001
Stem revision due to fracture within 2 years of the index procedure All uncemented stems All cemented stems 8.0 (4.5–14) < 0.001
Cup revision for any reason 5 most common uncemented cups 5 most common cemented cups 0.9 (0.6–1.1) 0.2
Cup revision due to aseptic loosening 5 most common uncemented cups 5 most common cemented cups 0.5 (0.3–0.8) 0.001
Stem revision for any reason 5 most common uncemented stems 5 most common cemented stems 0.5 (0.4–0.6) < 0.001