Skip to main content
. Author manuscript; available in PMC: 2011 Apr 6.
Published in final edited form as: J Pharm Biomed Anal. 2009 Nov 26;51(5):1120–1127. doi: 10.1016/j.jpba.2009.11.020

Table 2.

Intra-day and Inter-day precision and accuracy for matrine in MRM mode of UPLC–MS/MS analysis.

Analyte Concentration (ng/ml) Intra-day (n = 6)
Inter-day (n = 6)
Accuracy (bias, %) Precision (CV, %) Accuracy (bias, %) Precision (CV, %)
2.44 91.14 3.89 89.76 7.13
Matrine 78.0 106.44 4.62 103.08 3.31
1250 101.98 4.83 104.66 7.11