Skip to main content
. 2010 Mar 10;48(5):1782–1786. doi: 10.1128/JCM.02316-09

TABLE 5.

Numbers of errors and discrepancies between MICs obtained by commercial techniques and the CLSI reference procedure after 48 h of incubation, by antifungal agent

Antifungal agenta No. of errors or discrepancies between CLSI48H andb:
Vitek
Etest
SYOne
VME ME MiE SD NSD VME ME MiE SD NSD VME ME MiE SD NSD
AMB 0 0 0 8 0 10
5FC 1 0 14 0 1 23 0 0 16
FLZ 3 0 25 1 0 33 1 0 30
VOR 2 2 2 0 1 5 0 1 5
Total 6 2 41 0 0 1 2 61 0 8 1 1 51 0 10
a

AMB, amphotericin B; 5FC, flucytosine; FLC, fluconazole; VRC, voriconazole.

b

Vitek, Vitek 2 system technique; CLSI48H, CLSI reference procedure after 48 h of incubation; Etest, Etest method; SYOne, Sensititre YeastOne technique; VME, very major errors; ME, major errors; MiE, minor errors; SD, substantial discrepancies; NSD, nonsubstantial discrepancies.