. 2010 Mar 24;48(5):1661–1666. doi: 10.1128/JCM.00003-10

TABLE 1.

Relative sensitivity and specificity of the LightCycler MRSA advanced test and the BD GeneOhm MRSA assay compared to those of directly plated culturea

Site Relative sensitivity
Relative specificity
LightCycler MRSA advanced BD GeneOhm MRSA LightCycler MRSA advanced BD GeneOhm MRSA
1 91.4 (32/35; 76.9-98.2) 91.4 (32/35; 76.9-98.2) 97.5b (197/202; 94.3-99.2) 92.0b (184/200; 87.3-95.4)
2 95.8 (46/48; 85.7-99.5) 95.8 (46/48; 85.7-99.5) 95.9c (301/314; 93.0-97.8) 81.8c (257/314; 77.1-86.0)
3 100.0 (13/13; 75.3-100.0) 100.0 (13/13; 75.3-100.0) 97.6 (245/251; 94.9-99.1) 96.8 (243/251; 93.8-98.6)
4 95.0 (76/80; 87.7-98.6) 96.3 (77/80; 89.4-99.2) 93.9 (185/197; 89.6-96.8) 93.4 (183/196; 88.9-96.4)
5 100.0 (11/11; 71.5-100.0) 100.0 (11/11; 71.5-100.0) 96.8 (243/251; 93.8-98.6) 97.2 (243/250; 94.3-98.9)
Total 95.2 (178/187; 91.1-97.8) 95.7 (179/187; 91.7-98.1) 96.4c (1,171/1,215; 95.2-97.4) 91.7c (1,110/1,211; 90.0-93.2)
a

Results are shown as percentage (no. of results/total no. of samples; 95% CI [%]). There were 1,402 valid LightCycler MRSA advanced test results and 1,398 valid BD GeneOhm MRSA test results.

b

Difference between relative specificity results is P = 0.024.

c

Difference between relative specificity results is P < 0.001.