TABLE 2.

Relative sensitivity and specificity of the LightCycler MRSA advanced test, the BD GeneOhm MRSA assay, and directly plated culture compared to those of enrichment culture

Site Relative sensitivityf
Relative specificity
LightCycler MRSA advanced BD GeneOhm MRSA Directly plated culture LightCycler MRSA advanced BD GeneOhm MRSA Directly plated culture
1 83.3 (30/36; 67.2-93.6) 83.3 (30/36; 67.2-93.6) 86.1 (31/36; 70.5-95.3) 96.5a (194/201; 93.0-98.6) 91.0a,b (181/199; 86.1-94.6) 98.0b (197/201; 95.0-99.5)
2 87.9 (51/58; 76.7-95.0) 86.2 (50/58; 74.6-93.9) 81.0 (47/58; 68.6-90.1) 97.3c (289/297; 94.8-98.8) 82.5c,d (245/297; 77.7-86.6) 99.7d (296/297; 98.1-100.0)
3 81.3 (13/16; 54.4-96.0) 81.3 (13/16; 54.4-96.0) 81.3 (13/16; 54.4-96.0) 97.6b (242/248; 94.8-99.1) 96.8b (240/248; 93.7-98.6) 100.0b (248/248; 98.5-100.0)
4 97.5 (79/81; 91.4-99.7) 96.3 (78/81; 89.6-99.2) 95.1 (77/81; 87.8-98.6) 95.4 (187/196; 91.5-97.9) 93.8b (183/195; 89.5-96.8) 98.5b (193/196; 95.6-99.7)
5 78.6 (11/14; 49.2-95.3) 78.6 (11/14; 49.2-95.3) 78.6 (11/14; 49.2-95.3) 96.8b (240/248; 93.7-98.6) 97.2b (240/247; 94.2-98.9) 100.0b (248/248; 98.5-100.0)
Total 89.8 (184/205; 84.8-93.5) 88.8 (182/205; 83.6-92.8) 87.3 (179/205; 82.0-91.5) 96.8c,e (1,152/1,190; 95.6-97.7) 91.8c,d (1,089/1,186; 90.1-93.3) 99.3d,e (1,182/1,190; 98.7-99.7)
a

Difference between relative specificity results is P = 0.037.

b

Difference between relative specificity results is P < 0.05.

c

Difference between relative specificity results is P < 0.001.

d

Difference between relative specificity results is P < 0.001.

e

Difference between relative specificity results is P < 0.001.

f

Results are shown as percentage (no. of results/total no. of samples; 95% CI [%]).