Published in final edited form as: Med Image Comput Comput Assist Interv. 2007;10(Pt 2):553–561. doi: 10.1007/978-3-540-75759-7_67

Table 1.

Errors (in mm) were calculated as the distance between the predicted surface and twelve intraoperatively tracked points used for validation. Details about the deformation algorithm, which requires the preoperative locations of the sulci, can be found in [2].

mean max std
Total Deformation 4.49 7.24 2.03
Deformation Algorithm Error (Manually Segmented Sulci) 1.05 3.08 0.86
Deformation Algorithm Error (Semiautomatically Segmented Sulci) 0.58 1.01 0.22