Skip to main content
. 2010 Mar 29;5:15–23. doi: 10.4137/aci.s4431

Table 3.

Intra-day and inter-day precision (% RSD) and accuracy (% dev) for analysis of lumefantrine in human plasma.

Nominal, ng/mL Intra-day Inter-day

50 120 900 9000 20 120 900 9000
Mean, ng/mL 48.9–57.5 123–131 890–957 9082–9555 54.1 126 921 9371
SD 0.83–2.55 2.30–8.37 10.6–32.6 221–449 3.4 5.3 36 343
RSD, % 1.4–5.2 1.8–6.7 1.1–3.7 2.3–4.9 6.3 4.2 3.9 3.7
% dev −2.2–15 2.3–9.3 −1.1–6.4 0.9–6.2 8.2 5.1 2.3 4.1
n 5 5 5 5 25 25 25 25