Skip to main content
. 2009 Mar 13;467(9):2274–2280. doi: 10.1007/s11999-009-0773-2

Table 3.

Comparison of studies measuring wear on AL/AL hip

Study Method Cup Wear head Cup and head
Boutin and Blanquaert [8] Retrieved implants (10 years) 6 μm/year 9 μm/year
Walter [36] Simulator 6 μm/1million cycles
Bos et al. [6] Retrieved implants (10 years) 5 μm/year 5 μm/year
Refior et al. [28] Simulator 1 μm/1million cycles
Nevelos et al. [26] Retrieved implants (10 years) 5 μm/year 5 μm/year
Lusty et al. [22] Retrieved implants (2 years) 0.04 mm3/year 0.1 mm3/year
Hernigou et al. [current study] Retrieved implants (20 years)
Radiograph measurement
5 μm/year 5 μm/year 12 μm/year