TABLE 4.

Proposed Metrics for Standardized Dedicated Breast PET/PEM Evaluation

Guideline Task
Initial task Define phantoms appropriate for PEM-size scanner
Tests common to WB PET scanners Spatial resolution
Sensitivity (counting rate per activity)
RC vs. object size
Percentage background variability
Counting rate limitations, and count-loss corrections
Additional PEM-specific tests Quantification consistency within the FOV
Random coincidences from outside FOV
Variance at the edge of the FOV
Image position-to-physical space calibration accuracy for biopsy guidance