Skip to main content
. 2010 May 14;10:28. doi: 10.1186/1471-2466-10-28

Table 4.

Coefficients of repeatability (CR) retested after one hour (n = 237), after five hours (n = 79) and after 3-10 days (n-76) (in % from initial value/absolute value)

Parameter Retest after one hour Retest after five hours Retest another day
FEV1 11.8 13.7 17.9
IC 18.9 21.3 22.7
FIV1 13.5 17.9 18.0
PIF 17.9 17.8/0.90 17.9/0.85
MIF50 20.4 21.0 19.2

FEV1 = Forced Expiratory Flow in one second; FIV1 = Forced Inspiratory Flow in One second; IC- Inspiratory Capacity; MIF50 = Maximal Inspiratory Flow at 50%; PIF = peak Inspiratory Flow.