Table 2.

pH and pO2 a) measured with dual-sensor membrane

Set-point Measured values method A Measured values method C Measured values method B Measured values method D
pH b) 5.53c) 5.37 (-0.16) 5.42 (-0.11)
7.16d) 7.07 (-0.09) 7.14 (-0.02)
7.60 d) 7.63 (+0.03) 7.64 (+0.04)
pO2e) 5.1 4.6 (-0.5) 4.6 (-0.5)
8.1 8.4 (+0.3) 8.2 (+0.1)
10.1 9.7 (-0.4) 9.7 (-0.4)
12.2 12.0 (-0.2) 12.5 (+0.3)
15.2 14.6 (-0.6) 15.5 (+0.3)
20.2 19.0 (-1.2) 20.7 (+0.5)
(a)

pO2 represents the oxygen partial pressure in kPa;

(b)

Oxygen partial pressure of 21 kPa corresponding to [O2] of 8.6 mg L-1;

c)

pH 5.53 solution composed of tris(hydroxymethyl) aminomethane and hydrochloride acid with 10 mM NaCl;

d)

B-R buffers;

e)

Measurements carried out in pH 7.5 B-R buffer.