Skip to main content
. 2008 Nov;28(5):1233–1244. doi: 10.1002/jmri.21583

Table 2.

Maximum Temprature Changes for Different Scanners,RF Coils and Electrode Arragement.

Scanner[coil] SAR(W/Kg) head/body Tail termination ΔT°C
Depth Grid
GE 1.5T [Head coil] 2.4/0.2 open circuit <0.1 0.7
short circuit 2.5 1.7
repeated with depth electrode tips touching open circuit <0.1 0.7
short circuit 1.9 1.7
GE 3T[Head coil] 2.5/0.2 open circuit 0.3 2.0
short circuit 0.5 3.9
Siemens 3T [Head coil] 2.4/0.2 open circuit 0.2 0.9
short circuit 1.6 1.7
Siemens 3T [Body coil] 1.2/0.5 open circuit 0.7 6.4
short circuit 0.5 0.7
2.0/0.8 short circuit 0.8 1.2