Dose rates at point – A and point – B computed for FW HDR applicator

Reference points Dose Rate (Gy/h) (Sources only) *CB/CM

BrachyTPS (CB) MCNP (CM)
Point - A 7.422 7.544 (± 0.0003%) 0.983
Point - B 2.509 2.633 (± 0.0002%) 0.953
*

CB/CM = Ratio of BrachyTPS calculated value to MCNP calculated value