Skip to main content
. 2010 Apr-Jun;35(2):88–99. doi: 10.4103/0971-6203.62202

Table 4.

Dose rates at point – A and point – B computed for BRIT LDR applicator

Applicator Source Position Without Applicator With Applicator


Dose Rate cGy/h at Dose Rate cGy/h at


PointA (2.0, 2.0, 0.0) PointB (5.0, 2.0, 0.0) PointA (2.0, 2.0, 0.0) PointB (5.0, 2.0, 0.0)




Brachy TPS MCNP Brachy TPS MCNP Brachy TPS MCNP Brachy TPS MCNP
Fundus(S1) 28.0 28.982 9.74 10.213 27.6 28.574 9.59 10.149
Long Straight IUT Middle (S2) 46.2 47.954 8.66 8.838 45.7 46.745 8.58 8.777
Lower (S3) 50.9 50.857 9.16 8.920 49.1 50.510 8.96 8.907
Total 125.1 127.794 27.56 27.970 122.4 125.829 27.13 27.833
Right (S4) 18.9 18.544 9.51 9.306 18.6 18.295 9.32 9.253
Vaginal ovoids Left (S5) 9.51 8.876 4.16 4.063 9.32 8.530 4.05 4.005
Total 28.41 27.421 13.67 13.369 27.92 26.824 13.37 13.259
Long IUT & Vaginal Ovoids 153.51 155.214 41.23 41.339 150.32 152.654 40.50 41.091