Statistical Analysis of Manual Drawing and Automated Drusen Segmentation Results*

Photo Identification No. Eye Manual Drawing, % Automated, % False Positives, % False Negatives, % Sensitivity Specificity
1 Right 11.4 13.9 5.7 3.3 0.71 0.59
2a Left 33.5 29.5 6.5 10.5 0.69 0.78
2b Right 30.5 31.3 8.7 7.9 0.74 0.84
3 Right 50.8 39.8 6.8 17.30 0.66 0.83
4a Left 38.0 32.1 10.1 15.7 0.59 0.69
4b Right 36.5 36.9 11.2 10.5 0.71 0.70
5 Right 17.8 16.4 5.2 6.2 0.65 0.69
6 Left 45.2 37.2 5.3 13.3 0.71 0.86
7 Left 7.2 3.1 0.1 4.2 0.42 0.98
8a Left 36.6 28.6 5.5 13.4 0.63 0.81
8b Right 43.0 30.4 2.5 15.2 0.65 0.92
9 Left 43.2 35.1 4.7 12.9 0.70 0.87
10 Left 23.3 29.9 14.1 7.5 0.68 0.53
11 Left 27.6 34.8 1.8 8.9 0.74 0.94
12 Left 48.1 43.2 1.7 6.7 0.86 0.96
13 Right 34.6 34.1 6.5 7.0 0.80 0.81
14 Left 22.3 24.7 6.5 4.1 0.81 0.74
15 Left 41.1 32.3 1.1 9.9 0.76 0.97
16 Right 32.4 36.1 7.0 8.2 0.75 0.78
17 Right 2.5 1.9 0.4 1.1 0.58 0.78
*

Percentages refer to drusen area as a percentage of the 3000-μm-diameter circle. False positives, false negatives, sensitivity, and specificity are calculated on a pixel-by-pixel basis for the automated segmentation compared with the expert manual drawing.