. 2010 Apr 28;17(6):966–968. doi: 10.1128/CVI.00086-10

TABLE 1.

Comparison of the Bio-Rad BioPlex Syphilis IgG assay and the Trep-Chek IgG EIA using 1,008 prospective serum samples

Bio-Rad BioPlex Syphilis IgG assay result No. of Trep-Chek IgG EIA results
Positive Negative Total
Positive 77 14a 91
Negative 1b 916 917
Total 78 930 1,008
a

Eight out of these 14 samples were positive by TP-PA.

b

This sample became negative by EIA upon repeat testing.