TABLE 3.

Comparison of PCR/microneutralization and Luminex HAdV assay results obtained at NHRCa

HAdV serotype No. of samples with indicated results in the two assays
Sensitivity (%) Specificity (%)
W+ N+ W− N+ W+ N− W− N−
3 9 1 0 94 90 100
4 19 0 2b,c 83 100 98
7 10b 0 0 94 100 100
14 19c 0 0 85 100 100
21 17 3 0 84 85 100
a

Comparison of the results obtained by PCR/microneutralization and with the Luminex HAdV assay. W, result (+ or −) by the multiplexed Luminex HAdV assay; N, result (+ or −) by PCR/microneutralization. Sensitivity and specificity were based on true positives and true negative, defined as specimens giving positive or negative results according to the PCR/microneutralization assay. The sensitivity and specificity were calculated according to the methods reported in Cumitech 31 (9).

b

One HAdV-7 sample was positive for both HAdV-4 and HAdV-7.

c

One HAdV-14 sample was positive for both HAdV-14 and HAdV-4.