Table 2.
Distribution statistics for the experimental phantom maps of the factor F found on the right side of Fig. 11. The mean of F, standard deviation of F, and the scv value are given for 30°, 4.096 ms maximum-bandwidth block (BLK), BIR-4, and optimized composite (OPT) pulses. In terms of the width of the distribution of F, the optimized pulse performs slightly better that BIR-4 pulse when ΔB0 variations are minimal. When no B0 shimming is applied, the optimized pulse demonstrates superior flip-angle uniformity.
| ΔB0 | BLK | BIR-4 | OPT |
|---|---|---|---|
| SHIM | 0.69 ± 0.16(±23%) | 1.00 ± 0.09(±9%) | 1.03 ± 0.05(±5%) |
| NO SHIM | 0.69 ± 0.16(±23%) | 0.83 ± 0.20(±24%) | 1.04 ± 0.06(±6%) |