Table 1.
CID efficiencies of fibrinopeptide A (2+) and neurotensin 3+ inside the segmented quadrupole (SQ)
| Segmented quadrupole | Conductance limit orifice | ||||||
|---|---|---|---|---|---|---|---|
| V | Ec | Ef | ECID | V | Ec | Ef | ECID |
| fibrinopeptide A 2+ | |||||||
| 74 | 0.75 | 0.37 | 0.27 | 44 | 0.10 | 0.37 | 0.04 |
| 78 | 0.74 | 0.47 | 0.35 | 50 | 0.08 | 0.67 | 0.05 |
| 80 | 0.73 | 0.55 | 0.40 | 54 | 0.07 | 0.84 | 0.06 |
| 84 | 0.67 | 0.64 | 0.43 | 60 | 0.10 | 0.96 | 0.09 |
| 90 | 0.62 | 0.84 | 0.52 | 64 | 0.08 | 0.98 | 0.08 |
| 94 | 0.64 | 0.92 | 0.59 | 70 | 0.09 | 1.00 | 0.09 |
| 100 | 0.63 | 0.97 | 0.61 | ||||
| 110 | 0.60 | 1.00 | 0.60 | ||||
| neurotensin 3+ | |||||||
| 80 | 0.64 | 0.94 | 0.60 | 60 | 0.02 | 0.94 | 0.02 |
V: collision voltage gradient, V/mm; Ec: collection efficiency; Ef: fragmentation efficiency; ECID: CID efficiency.