Skip to main content
. 2010 Jun 10;5:52. doi: 10.1186/1748-717X-5-52

Figure 2.

Figure 2

Isocenter placement errors (in cm) relative to DRR on during-treatment EPIs (gray circles; n = 530 fractions), along a): S-I and A-P axes, and b): S-I and L-R axes. Outer box shows PTV margins used in the study; inner ellipse shows 95% confidence intervals for CTV coverage in each direction.