Skip to main content
. 2010 Jul 7;92(7):1652–1660. doi: 10.2106/JBJS.I.01111

Fig. 4.

Fig. 4

Torsion test results after tibial harvest and implant removal at postoperative week 9 (mean and standard deviation). Compared with the locked plating (LP) group, tibiae in the far cortical locking (FCL) group were not significantly stiffer (A), but they healed to be 54% stronger (B), and they sustained 156% more energy before fracturing (C).