Skip to main content
. 2010 Apr 30;76(13):4143–4150. doi: 10.1128/AEM.03059-09

TABLE 3.

Detection limits for botulinum neurotoxin type A, B, E, and F complexes and E. coli O157 in various matrices using the AICC system

Matrix Detection limit of AICC assay fora:
Botulinum neurotoxin type A (ng) Botulinum neurotoxin type B (ng) Botulinum neurotoxin type E (ng) Botulinum neurotoxin type F (ng) E. coli O157 (cells)
Semiskim milk 0.5 50 5 50 500
Skim milk 0.5 50 5 50 500
Apple juice 500
Grape juice with peach juice 0.5 50 50 500
Bottled water (type 1) 0.5 50 5 50 500
Bottled water (type 2) 0.5 50 5 50 500
a

The volume for each test was 500 μl. −, no detection.