Skip to main content
. 2010 May 19;48(7):2387–2395. doi: 10.1128/JCM.00220-10

TABLE 2.

Validation of reference real-time respiratory virus PCR assaysa

Multiplex assay Virus target(s) No. of specimens with pos PCR VL result/no. tested (% PCR pos)
FA staining pos FA staining neg, culture pos FA staining neg, culture neg
1b Influenza A virus 30/30 (100) 30/30 (100) 1/152 (0.7)
1b Influenza B virus 31/31 (100) 27/30 (90) 0/152
3 Respiratory syncytial virus 30/30 (100) 9/9 (100) 6/39 (15)
4 Parainfluenza virus 16/16 (100) 8/11 (73) 0/10
5 Adenovirus 3/3 (100) 26/26 (100) 0/33
6 Human metapneumovirus 16/16 (100) NAc 7/104 (7)c
6 Rhinovirus NAd 27/27 (100)d 26/66 (39)d
a

NA, not applicable; pos, positive; neg, negative.

b

The influenza virus multiplex assay detects both influenza A and B viruses and can distinguish between them with different probe labels.

c

No cultures were performed for human metapneumovirus. Each of the seven specimens that were positive by reference PCR was also positive by other PCR assays (1).

d

No FA staining was performed.