. 2010 May 1;4(3):522–531. doi: 10.1177/193229681000400304

Table 2.

Studies of Dose Accuracy of Insulin Pens with Comparable End Points

Original FlexPena Improved FlexPen HumaPen LUXURA SoloSTAR OptiClik Reference
Deviation from 10-IU dose (mean) 1.64 ± 0.84% NAb 1.10 ± 0.20% 2.61 ± 0.92% 4.78 ± 3.31% 59
NA 1.63 ± 0.84%c NA 2.11 ± 0.92% NA 64
Deviation from 30-IU dose (mean) 0.83 ± 0.26% NA 0.62 ± 0.19% 1.70 ± 0.84% 2.97 ± 2.48% 59
NA 1.23 ± 0.76%d NA 1.54 ± 0.84% NA 64
Mean delivered dose of 5 IU 5.07 ± 0.15 NA NA 5.03 ± 0.21 NA 61
4.95 ± 0.19 NA NA 4.86 ± 0.39 NA 62
Mean delivered dose of 10 IU 9.87 ± 0.16 NA NA 9.83 ± 0.14 NA 61
9.61 ± 0.27 NA NA 9.27 ± 0.52 NA 62
Mean delivered dose of 30 IU 29.70 ± 0.38 NA NA 29.45 ± 0.25 NA 61
29.70 ± 0.34 NA NA 28.73 ± 0.47 NA 62
a

No longer available commercially.

b

Not assessed.

c

p < 0.001

d

p < 0.05 for improved FlexPen vs SoloSTAR.