Skip to main content
. 2010 May-Jun;63(3):212–224. doi: 10.4212/cjhp.v63i3.917

Table 4.

Stability of Ceftriaxone during Storage in Accufuser Elastomeric Infusion Device, Expressed as Percentage of Original Concentration Remaining (Mean ± SD)*

Study Day Nominal 5 mg/mL
Nominal 40 mg/mL
D5W
NS
D5W
NS
23°C 4°C 23°C 4°C 23°C 4°C 23°C 4°C
Observed initial concentration (mg/L) 5.65 ± 0.47 5.14 ± 0.29 5.10 ± 0.42 5.77 ± 0.85 42.14 ± 3.83 42.69 ± 0.22 42.44 ± 2.04 45.45 ± 1.99
1 97.0 ± 2.2 101.0 ± 4.3 98.7 ± 3.8 104.7 ± 14.5 97.8 ± 9.7 97.2 ± 0.7 96.7 ± 1.6 97.94 ± 3.2
2 94.5 ± 3.6 97.4 ± 1.6 93.3 ± 6.1 104.8 ± 11.9 95.8 ± 0.3 99.3 ± 4.6 94.5 ± 4.6 98.2 ± 3.1
4 92.6 ± 4.6 96.4 ± 0.7 92.6 ± 7.6 105.8 ± 15.6 95.6 ± 0.8 97.3 ± 5.1 95.1 ± 3.4 98.7 ± 5.8
7 89.2 ± 0.6 97.6 ± 8.2 91.0 ± 5.7 104.6 ± 12.8 87.0 ± 1.0 96.3 ± 2.0 88.7 ± 2.1 95.7 ± 2.4
14 86.1 ± 10.7 97.3 ± 4.1 84.5 ± 4.6 103.6 ± 13.1 69.7 ± 1.2 95.8 ± 2.4 77.7 ± 2.6 95.5 ± 3.0
21 94.3 ± 3.6 97.4 ± 13.4 91.0 ± 1.9 89.8 ± 3.7

Degradation rate (%/day) –1.09 –0.22 –0.99 –0.22 –2.15 –0.35 –1.51 –0.41
  T–90% (days) 7.4 41.9 8.1 65.7 5.2 26.1 6.1 23.1

Fastest degradation rate§
Lower limit of 95% CI (%/day) –1.37 –0.43 –1.23 –0.60 –2.39 –0.52 –1.69 –0.58
  T–90% (days) 7.3 23.5 8.1 16.7 4.2 19.4 5.9 17.4

Estimated % remaining with storage
  13 days at 4°C 94.47 92.23 93.30 92.53
  For 1 additional day at 23°C 93.10 91.00 90.91 90.84

CI = confidence interval, D5W = dextrose 5% in water, NS = normal saline (0.9% sodium chloride), SD = standard deviation.

*

Each value was based on duplicate determination of 3 samples. The percent remaining was based on the observed concentration on day 0 (100%).

The degradation rate was determined by linear regression of the percent remaining on each study day.

T–90% is the time for the concentration to decline by 10% (i.e., time to reach 90% of initial concentration), as determined from the degradation rate.

§

Fastest degradation rate and T–90% were based on the lower 95% confidence limit of the regression-determined slope.

Using the fastest observed degradation rates determined by regression (95% CI), calculations were completed for a solution stored in an Accufuser elastomeric infusion device for 13 days, followed by storage for an additional 24 h at room temperature (23ºC).