Skip to main content
. 2010 Jun 16;37(7):3587–3594. doi: 10.1118/1.3447724

Table 2.

Peripheral dose (mGy) at the locations of interest in the Rando phantom. The doses reported are the average of the readings from two TLDs (uncertainty 5%); in cases where the values were different for more than 7%, both readings were reported. Abbreviations: TLDs=thermoluminescent dosimeters; CC=cranio-caudal; LINAC=linear accelerator; mMLC=micromultileaf collimator; CK=CyberKnife; R=right; nv=nerve; L=left.

TLDs anatomical position CC distance from the target (cm) LINAC-mMLC (mGy) LINAC-cone (mGy) CK preshileding (mGy) CK postshielding (mGy) TomoTherapy (mGy) Gamma Knife (mGy)
Target ant 0 8200 9695∕9199 7900 6908 5322 8314∕7587
Target post 0 7500 6915∕6234 6800 7425 3588 7603∕7427
R optic nv 0 480 533∕414 870 867∕464 1772∕1940 366∕290
L optic nv 0 870 1106∕1119 340 369 2028 454∕519
Thyroid 15.5 45.0 62.8 55.0 20.9 6.9 13.2
Sternum 30.5 10.0 18.0 27.5 9.1 2.3 4.1
Upper lung 43.0 4.6 8.7 17.5 7.7 1.7 1.9
Lower lung 53.0 3.0 5.8 15.0 8.3 1.4 0.98
Ovaries 75.5 1.3 2.6 10 5.8 1.6 0.36
Testes 80.5 1.1 2.2 10 5.8 1.6 0.28