. 2010 Jun 16;37(7):3587–3594. doi: 10.1118/1.3447724

Table 3.

Comparison between the peripheral dose values measured in the CyberKnife treatments before and after the supplemental shielding was installed.

Distance from the target (cm) CyberKnife preshielding dose (mGy) CyberKnife postshielding dose (mGy) Dose reduction after shielding application (%)
30.5 27.5 9.1 66.9
43 17.5 7.7 56.0
53 15 8.3 44.7
75.5 10 5.8 42.0
80 10 5.8 42.0