Skip to main content
. 2000 Dec 8;5(1):24–30. doi: 10.1186/cc974

Table 3.

Maximum and minimum tracheal temperatures

Tracheal temperatures (°C)

Inspired gas-conditioning
device Maximum Minimum Room temperature (°C) Body temperature (°C)
HH
 45 min 35.7 ± 1.0 34.3 ± 1.0 23.6 ± 0.9 37.8 ± 1.0
(35.1-36.5) (33.5-35.1) (22.5-24.9) (36.4-39.1)
 24 h 35.9 ± 0.9 34 ± 1.0 23.3 ± 1.0 37.4 ± 0.9
(35.2-37.1) (33.2-34.9) (22.6-25.0) (36.3-38.4)
Hydrophobic HME
 45 min 35.4 ± 1.2 32.0 ± 1.2 23.6 ± 0.9 37.9 ± 1.0
(33.4-37.3) (30.4-33.2) (22.6-25.0) (36.1-39.0)
 6 h 35.5 ± 0.8 32.6 ± 0.8 23.6 ± 0.9 37.9 ± 1.0
(34.4-36.7) (31.3-33.4) (22.0-24.9) (36.0-39.4)
 24 h 35.3 ± 0.7 32.4 ± 1.1 23.6 ± 1.0 37.3 ± 0.8
(34.4-36.1) (30.5-32.9) (21.9-24.2) (36.3-38.4)
Hygroscopic HME
 45 min 36.3 ± 0.8 34.2 ± 0.9 23.6 ± 1.0 37.4 ± 1.1
(35.0-36.9) (33.1-34.8) (21.9-24.6) (36.1-39.5)
 6 h 36.2 ± 0.9 34.2 ± 0.7 23.4 ± 1.5 37.9 ± 1.0
(35.0-37.3) (33.4-35.0) (20.8-24.2) (36.5-39.2
 24 h 36.2 ± 0.8 34.3 ± 1.2 23.0 ± 1.2 37.8 ± 0.8
(35.5-37.8) (32.5-37.7) (20.8-24.4) (36.5-38.8)

Room temperature was measured at the patients' bedsides. Body temperature was obtained in the oesophagus. Values are expressed as mean ± standard deviation (range).