Published in final edited form as: Magn Reson Med. 2010 Jun;63(6):1548–1556. doi: 10.1002/mrm.22329

Table 1.

CBF Measurements Using the Different Techniques (n = 20 Patients)

Method Tmaxthresh CBF (ml/100 g/min) CBF Ratio COV
xeCT (gold-standard) 35.8 ± 7.7
DSC 14.6 ± 6.3* 0.43 ± 0.21 47%
ASL 31.4 ± 9.9 0.91 ± 0.31 34%
CAD-CBF 2 sec 29.0 ± 10.1 0.83 ± 0.29 35%
3 sec 31.2 ± 9.8 0.90 ± 0.30 33%
4 sec 32.0 ± 10.1 0.93 ± 0.32 34%
5 sec 32.0 ± 10.0 0.93 ± 0.31 33%
6 sec 31.8 ± 10.0 0.92 ± 0.31 32%
31.4 ± 9.9 0.92 ± 0.31 32%
DSC scaled by white matter ROI of 22 mL/100 g/min 43.4 ± 7.4 1.29 ± 0.46 36%

All values mean ± standard deviation. Tmaxthresh: Tmax threshold for including voxel in the calculation of the correction factor. CBF ratio = mean MR-based CBF value divided by the mean xeCT CBF value (i.e., for no bias, CBF ratio = 1.0). COV = SD normalized by the mean (for perfect precision, COV = 0%).

*

Significantly different from xeCT CBF (P < 0.01, paired t test).