Table 3.

Statistical analysis for the ionization chamber array measurements.

IC position Mean SD abs. mean abs SD minimum maximum
target volume 0.3 1.5 1.2 0.9 -2.7 3.7

left penumbra -13.3 2.3 13.3 2.3 -15.8 -10.5
right penumbra 7.8 0.1 7.8 0.1 7.8 7.9
distal edge -1.5 5.1 4.1 3.1 -8.9 3.3
position A 0.8 1.6 1.4 1.0 -2.5 3.7
position B -0.2 1.3 1.0 0.8 -2.7 2.3

All values denote the relative dosimetric deviation between motion compensated and the stationary reference measurement in %. Position A refers to chamber positions indicated by solid lines, position B to chamber positions indicated by thin lines in fig. 1c.