Table II.
Parameter | Required Setting |
---|---|
* Pacing mode | DDD or DDDR |
# LRL (lower rate limit) | 60 |
ATR (atrial tachy response) | ON at 170 beats/min |
Rate hysteresis | OFF |
Dynamic atrioventricular (AV) delay | OFF |
† AV delay | Fixed arm: 120 ms SmartDelay arm: per algorithm Echocardiogram arm: per iterative method |
‡ Left ventricular (LV) offset | Fixed arm: 0 SmartDelay arm: 0 Echocardiogram arm: recommend 0 |
Set after the baseline echocardiogram (device programmed to VVI-40-RV) at the post-implant.
Set after the baseline echocardiogram (device programmed to VVI-40-RV) at the post-implant.
Only patients randomized to the SmartDelay arm will have their AV delay settings re-optimized at the 3-month follow-up visit.
For the echocardiogram arm only, after the AV delay is determined, an LV offset other than 0 may be programmed.