Table II.

Required Device Programming

Parameter Required Setting
* Pacing mode DDD or DDDR
# LRL (lower rate limit) 60
ATR (atrial tachy response) ON at 170 beats/min
Rate hysteresis OFF
Dynamic atrioventricular (AV) delay OFF
AV delay Fixed arm: 120 ms
SmartDelay arm: per algorithm
Echocardiogram arm: per iterative method
Left ventricular (LV) offset Fixed arm: 0
SmartDelay arm: 0
Echocardiogram arm: recommend 0
*

Set after the baseline echocardiogram (device programmed to VVI-40-RV) at the post-implant.

#

Set after the baseline echocardiogram (device programmed to VVI-40-RV) at the post-implant.

Only patients randomized to the SmartDelay arm will have their AV delay settings re-optimized at the 3-month follow-up visit.

For the echocardiogram arm only, after the AV delay is determined, an LV offset other than 0 may be programmed.