Table 5.
Four Written User Scenarios with Distractions and Errors
Scenario | Distractions | Tasks to complete (pathway) | |
---|---|---|---|
1 | Normal entry | Someone will open the door and talk on the phone | Enter data as written on cue card, press the “Submit” button, receive bolus guide recommendation |
2 | Out of range value | None | Enter data as written on cue card, respond to messages by pressing “OK,” take corrective action, enter corrected values (to be provided on cue card), press the “Submit” button, receive bolus guide recommendation |
3 | BG measurement time error (>10min) | Radio | Enter data as written on cue card, respond to messages, take corrective action, enter new time (to be provided on cue card), press the “Submit” button, receive bolus guide recommendation |
4 | Previous bolus list entry error | Telephone call | Enter data as written on cue card, respond to messages, take corrective action, enter missing time (to be provided on cue card), press the “Submit” button, receive bolus guide recommendation |