Table 2.
Precision and accuracy for a HPLC assay of naltrexone from rat plasma (n = 6)
Theoretical con- centration, ng/ml |
Precision, % RSD |
Accuracy, % |
||
---|---|---|---|---|
intra-batch | inter-batch | intra-batch | inter-batch | |
10 | 12.02 | 17.19 | 90.6 | 86.48 |
100 | 7.75 | 4.11 | 92.94 | 96.66 |
500 | 2.37 | 0.14 | 97.90 | 99.88 |