MOSFET sensitivity for varying fraction sizes

Fraction Size (cGy) Dose per ‘beam’ (cGy) Sensitivity (mV/cGy)
203 29 2.63
252 36 2.63
301 43 2.65
350 50 2.65
497 71 2.63
749 107 2.61
1001 143 2.62