Table 1.

Inter- and Intra-day precision and accuracy of topotecan lactone in whole blood.

Nominal concentration (ng/ml) Mean (SD) Measured concentration (ng/ml) Accuracy (%) Within-day precision (%RSD) Between-day precision (%RSD)
7.0 7.2 (0.3) 102.3 3.2 4.1
30.0 28.1 (2.1) 93.7 2.9 7.6
80.0 75.8 (5.1) 94.8 2.6 6.7

RSD. Relative standard deviation (SD*100/mean)