. 2010 Jul 14;37(8):4102–4109. doi: 10.1118/1.3426000

Table 2.

Techniques used on each make and model of scanner to obtain 20 repeated free-in-air CTDI measurements using a 10 cm pencil ionization chamber suspended at isocenter. All scans were performed in axial mode; nominal beam width can be calculated by multiplying the number of channels by the channel width.

Scanner make and model kVp mA Exposure time (s) No. of channels Channel width (mm) Bowtie
GE VCT 120 350 1 64 0.625 Large
GE LightSpeed 16 120 250 1 16 1.25 Large
Siemens Sensation 64 120 350 1 24 1.2 Large