Table 3.

kVp, type of bowtie, and collimation variations used on each make and model of scanner to gather free-in-air CTDI measurements; axial scan mode and 1 s exposure times were used for all techniques. A total of 21, 12, and seven techniques were tested on each of the three GE VCT, GE LightSpeed 16, and Siemens Sensation 64 units, respectively.

Scanner make and model kVp mA s No. of channels Channel width (mm) Bowtie
GE VCT 80, 100, 140 350 64 0.625 Small, medium, large
120 350 8, 16, 32, 64 0.625 Small, medium, large
GE LightSpeed 16 80 350 16 1.25 Small, large
100, 140 250 16 1.25 Small, large
120 250 4, 8, 16 1.25 Small, large
Siemens Sensation 64 80, 100, 120, 140 350 24 1.2 Large
120 350 32 0.6 Large
120 350 1, 2 5 Large