Skip to main content
. 2010 Jul 2;7:11. doi: 10.4103/1742-6413.65057

Table 1.

APTIMA AC2 Chlamydia (CT) and Gonorrhea (GC) Results - Endocervical Swab vs. SurePath (SP) Samplesa

SP CT positive (n = 167) SP CT negative (n = 131) SP CT Sensitivity/Specificity (%)
CT Results
 Swab CT Positive (n = 187) 167 20 Sensitivity 89.3 (95% CI 84.0-93.3)
 ≤ 25 yr (n = 125) 121 9 Sensitivity 93.1 (95% CI 87.2-96.8)
 > 25 yr (n = 62) 46 11 Sensitivity 80.7 (95% CI 68.1-90.0)
 Swab CT negative (n = 111) 0 111 Specificity 100.0 (95% CI 96.7-100.0)
GC Results
 Swab GC positive (n = 19) 19 0 Sensitivity 100.0 (95% CI 82.4-100.0)
 Swab GC negative (n = 281) 0 281 Specificity 100.0 (95% CI 98.7-100.0)
a

Two SP equivocal results, one endocervical swab CT positive, and one endocervical swab CT negative excluded