Skip to main content
. Author manuscript; available in PMC: 2011 Sep 1.
Published in final edited form as: J Thorac Cardiovasc Surg. 2010 Jun 18;140(3):529–536. doi: 10.1016/j.jtcvs.2010.04.037

Table 1.

Transition from flow stabilizer to pump with increased rotation.

RPM Pressure rise (mmHg) Maximum Wall Shear (Pa)* Power Loss (Watts)
No impeller −0.43 1.4 0.25
0 −0.05 7.6 0.03
500 −0.05 15.6 0.03
1000 −0.04 29.3 0.02
2500 0.02 98.7 −0.01
5000 0.03 130.8 −0.02

RPM: revolutions per minute; Pa: pascal,

*

location of maximum wall shear stress is on the TCPC vessel wall for case with no impeller and on impeller surface for cases with impeller