Change in cantilever deflection on exposure to neutravidin. In one
case, the silicon nitride surface of the cantilever was coated with
biotin and polyethylene glycol, whereas in the other case, the same
surface cantilever was coated with a silane layer. Specific binding of
neutravidin to biotin generated a negative deflection, indicating
generation of compressive stress on the top silicon nitride surface,
whereas exposure to silane produced no appreciable deflection signal.