Table 4.

Results of uniformity and contrast measurements for phantom experiments simulating low-dose rest part of MPS acquisition, including lateral and septal wall defect inserts. Each measurement represents a series of ten phantom acquisitions performed sequentially on DNM and S-SPECT (Ventri). Overall, notwithstanding the significantly shorter acquisition time of DNM, uniformity and defect contrast trade-off are comparable to S-SPECT

Male phantom configuration mean (SD) Female phantom configuration mean (SD)
DNM S-SPECT DNM S-SPECT
NSD, % 12.5 (1.0) 13.6 (1.35) 14.9 (0.98) 17.5 (2.55)
Lat EC, % 38.5 (4.75) 30.5 (8.5) 37.5 (6.35) 42.3 (6.05)
Sep EC, % 28.5 (4.1) 14.2 (5.55) 33.8 (5.8) 20.3 (10.55)

SD standard deviation, NSD normalized standard deviation, Lat lateral wall defect, EC effective contrast, Sep septal wall defect