Agreement between the Smart GBS LB assay and BD GeneOhm StrepB assaya

BD GeneOhm StrepB result No. of samples with indicated Smart GBS LB result
Positive Negative Total
Positive 66 4 70
Negative 30 206 236
Total 96 210 306
a

Positive result agreement was 68.8%, and negative result agreement was 98.1%.