Skip to main content
. 2010 Jul 7;48(9):3088–3094. doi: 10.1128/JCM.00785-10

TABLE 2.

Clinical performance characteristics of the RealTime Ready Influenza A/H1N1 Detection Seta

RealTime Ready Influenza A/H1N1 Detection Set No. of results by alternative RT-qPCR assays
Total no. of results % agreement (95% CI)
Pandemic (H1N1) 2009 virus positive Seasonal influenza A virus positive Negative
Pandemic (H1N1) 2009 virus positive 112 0 7 119 88, negatives (76-90)
Seasonal influenza A virus positive 2 75 13 90 99, seasonal influenza A virus (93-100)
Negative 0 1 149 150 98, pandemic (H1N1) 2009 virus (94-100)
Total 114 76 169 359
a

Alternative assays used were the CDC rRT-PCR assay for 2009 H1N1 influenza virus or the Focus Diagnostics Influenza A H1N1 (2009) real-time RT-PCR assay. CI, confidence interval.