Table 3.

Accuracy and precision of the approach for relative quantification using LC-MS and 18O labeling as determined by measuring three replicates of Inline graphic mixtures prepared at the heavy-to-light ratios noted

Expected Measured CV (%)
5′-[m5U]ΨCGp-3′
[M-H] (m/z 1293.16) 0.20 0.187 ± 0.0045 2.38
1.00 1.11 ± 0.029 2.61
5.00 5.13 ± 0.0395 0.769
10.00 9.16 ± 0.341 3.72
[M-2H]2- (m/z 646.074) 0.20 0.141 ± 0.0266 18.8
1.00 1.09 ± 0.0249 2.28
5.00 5.50 ± 0.142 2.58
10.00 9.76 ± 0.370 3.79
5′-A[s4U]UAGp-3′
[M-H] (m/z 1648.18) 0.20 0.113 ± 0.0146 13.0
1.00 0.991 ± 0.130 13.1
5.00 4.56 ± 0.285 6.24
10.00 10.05 ± 0.919 9.14
[M-2H]2- (m/z 823.587) 0.20 0.117 ± 0.00639 5.48
1.00 0.959 ± 0.0525 5.47
5.00 4.96 ± 0.336 6.78
10.00 9.21 ± 0.341 3.70

Quantification was done using the RNase T1 digestion products 5′-[m5U]ΨCGp-3′ and 5′-A[s4U]UAGp-3′, which were found to yield linear and consistent results. CV, coefficients of variation.