Table 3.

Effect of 1H decoupling on linewidths (reported in Hz) and SNR.

Line widths* SNR#
undecoupled decoupled difference Sdecoupled/Sundecoupled
[1-13C]pyr [1-13C]lac [1-13C]pyr [1-13C]lac [1-13C]pyr [1-13C]lac [1-13C]pyr [1-13C]lac
7.2 ± 1.1 12.8 ± 0.9 6.0 ± 1.0 6.4 ± 0.6 1.2 ± 0.3 6.3 ± 0.6 1.12 1.76
[2-13C]pyr [2-13C]lac [2-13C]pyr [2-13C]lac [2-13C]pyr [2-13C]lac [2-13C]pyr [2-13C]lac
22 ± 4 10 ± 1 11 ± 1 5 ± 1 11 ± 2 4 ± 1 1.50 2.36
*

obtained experimentally using LASER sequence

#

obtained using simulations