Table 3.
Limits of quantitation (LOQ), linear dynamic range, and intra- and inter-day precision in the determination of vitamin D metabolites from plasma
Component | LOQ pg/mL (plasma) | Linearity (pg/μL) | r 2 | Intra-day (%CV) | Inter-day (%CV) |
---|---|---|---|---|---|
24,25-(OH)2 D3 | n/aa | 0.1–300 | 0.9995 | 4.1 | 6.8 |
1,25-(OH)2 D3 | 20 | 0.2–50 | 0.9990 | 3.9 | 6.1 |
1,25-(OH)2 D2 | 10 | 0.1–50 | 0.9992 | 2.2 | 6.2 |
25-OH D3 | 10 | 2.5–300 | 0.9999 | 3.8 | 5.8 |
25-OH D2 | 10 | 2.5–300 | 0.9996 | 1.6 | 3.7 |
No LOD was given because of the lack of a commercial deuterated standard. 24,25(OH) D3 and 1,25(OH) D3 were similar, and the LOQ is estimated to be similar to that of 1,25(OH) D3
a24,25-(OH)2 D3 was quantified using D6-25-OH D2 as internal standard