Published in final edited form as: Acad Radiol. 2009 Jun;16(6):689–699. doi: 10.1016/j.acra.2008.12.024

Table 5.

Comparability of lung density measures from cardiac CT scans on EBT vs. MDCT scanners

n=10 Median value from EBT cardiac scan (interquartile range) Median value from MDCT* cardiac scan (interquartile range) Spearman's correlation coefficient Mean difference (95% CI) 95% limits of agreement
%emphysema-910 HU 10.8 (7.35, 16.9) 10.9 (6.94, 18.5) 0.94 -0.87 (-2.20, 0.46) -5.07, 3.34
Alpha-910 HU 1.38 (1.19, 1.59) 1.45 (1.26, 1.61) 0.83 -0.04 (-0.10, 0.02) -0.23, 0.14
Apical-basilar difference in %Emphysema-910 HU 0.12 (-4.35, 1.63) -3.27 (-6.58, -1.00) 0.75 2.04 (0.05, 4.03) -4.25, 8.33
Apical-basilar difference in alpha-910 HU 0.13 (0.09, 0.35) 0.11 (0.07, 0.13) 0.22 0.09 (-0.03, 0.22) -0.30, 0.49
*

Siemens 64-slice multidetector CT scanner

Mean difference calculated as EBT value – MDCT value