Published in final edited form as: Magn Reson Med. 2009 Jun;61(6):1374–1387. doi: 10.1002/mrm.21960

Table 1.

Test-retest repeatability, COVF (standard deviation of perfusion measurement as a percentage of mean flow), for each combination of breathing and background suppression strategies

Between Session Repeatability
Breath-Hold Timed Breathing Free Breathing

Heavy Bckgrd Supp 7 % 23 % 20 %

Moderate Bckgrd Supp 12 % 15 % 19 %

None 18 %

In-Session Repeatability
Heavy Bckgrd Supp 8 % *

COVF is significantly lower:

*

(10% level, or better, for the F-test) compared to all other measurements except those for Breath-Hold,

5% level, or better, compared to all other between-session measurements.